Laser-induced fluorescence measurements of dual plumes and comparison of laser-induced fluorescence and conductivity probe measurements

Loading...
Thumbnail Image
Author(s)
Chrzan, Joseph Coleman
Editor(s)
Associated Organization(s)
Series
Supplementary to:
Abstract
The laser-induced fluorescence (LIF) technique is used to visualize and quantify the concentration field around a conductivity probe. The LIF data are compared to the signal collected by the conductivity probe. The objective is to compare the signal of the contact-sensor to the "ground-truth" measurement of the LIF data. Detailed comparison of the temporal response and the peak detection are presented. In addition, a proof-of-concept of a two-color LIF technique is presented using Rhodamine 6G paired with an Argon-ion laser and simultaneously Oxazine 725 paired with a Krypton-ion laser. Optical filters on two digital cameras isolate the emitted light from these respective laser/dye combinations. The objective is to provide detailed quantitation of two over-lapping (non-reactive)chemical plumes.