Quantitative Evaluation of the Microsoft Kinect for Use in an Upper Extremity Virtual Rehabilitation Environment

Loading...
Thumbnail Image
Author(s)
Nixon, Mason
Chen, Yu-ping
Advisor(s)
Editor(s)
Series
Supplementary to:
Abstract
Low cost depth sensors could potentially allow for home-based care and rehabilitation using virtual systems. Currently, no publicly available and peer-reviewed assessment has been made on the accuracy of joint position data determined by the Microsoft KinectTM for assessment of upper extremity movements. We devised and validated clinically-based angle classifications for random arm movements in 3D-space, specifically, the shoulder joint flexion/extension angle, shoulder joint abduction/adduction angle, and 3-dimensional shoulder joint angle of 19 subjects at a distance of 2.0m using an eight camera Vicon Motion Capture system. Results show an average absolute error of these angle measurements not exceeding 10.0%.
Sponsor
Date
2013
Extent
Resource Type
Text
Resource Subtype
Proceedings
Post-print
Rights Statement
Rights URI