Title:
Assessment of the ISO Impact Damping Test

Thumbnail Image
Author(s)
Sprigle, Stephen
Chung, Bummo
Meyer, Tobias
Authors
Advisor(s)
Advisor(s)
Editor(s)
Associated Organization(s)
Organizational Unit
Series
Supplementary to
Abstract
The ISO impact damping test characterizes wheelchair cushion abilities to reduce impact loading on issues and to help maintain postural stability, reporting the number of rebounds greater than 10% of the peak impact acceleration and the ratio of the 2nd to 1st rebound. Based upon our analysis, 3 critical issues have been identified. 1) Impact magnitude should be part of the analysis. 2) Impact should be used instead rebound accelerations. 3) Oscillation from impact is not a simple 2nd order damped harmonic, instead several natural frequencies are embedded in the damped oscillation. In conclusion, ISO should utilize impact accelerations and more complex analysis to characterize the damping properties of wheelchair cushions.
Sponsor
Date Issued
2012
Extent
Resource Type
Text
Resource Subtype
Poster
Rights Statement
Rights URI