Enhancing the image quality of digital breast tomosynthesis

Loading...
Thumbnail Image
Author(s)
Feng, Si
Advisor(s)
Sechopoulos, Ioannis
Editor(s)
Associated Organization(s)
Organizational Unit
Wallace H. Coulter Department of Biomedical Engineering
The joint Georgia Tech and Emory department was established in 1997
Organizational Unit
Supplementary to:
Abstract
A novel imaging technology, digital breast tomosynthesis (DBT), is a technique that overcomes the tissue superposition limitation of conventional mammography by acquiring a limited number of X-ray projections from a narrow angular range, and combining these projections to reconstruct a pseudo-3D image. The emergence of DBT as a potential replacement or adjunct to mammographic screening mandates that solutions be found to two of its major limitations, namely X-ray scatter and mono-energetic reconstruction methods. A multi-faceted software-based approach to enhance the image quality of DBT imaging has the potential to increase the sensitivity and specificity of breast cancer detection and diagnosis. A scatter correction (SC) algorithm and a spectral reconstruction (SR) algorithm are both ready for implementation and clinical evaluation in a DBT system and exhibit the potential to improve image quality. A principal component analysis (PCA) based model of breast shape and a PCA model of X-ray scatter optimize the SC algorithm for the clinical realm. In addition, a comprehensive dosimetric characterization of a FDA approved DBT system has also been performed, and the feasibility of a new dual-spectrum, single-acquisition DBT imaging technique has also been evaluated.
Sponsor
Date
2013-05-14
Extent
Resource Type
Text
Resource Subtype
Dissertation
Rights Statement
Rights URI